JX7100B Portable Anesthesia Machine
Product Parameters
- Number:
- PAM003
- Specifications:
- No Attribute
- Type:
- No Attribute
- Attribute:
- No Attribute
Introduction
It is an excellent portable anesthesia machine
- Configuration
- Function Introduction
- Instructions
GAS Supply
·Gas source:
3 gas, O2,N2O,Air
·Gas source pressure gauge:
3 gas, O2,N2O,Air
·Output pressure of regulator: 280kpa~300kpa
·Security valve release pressure:≦600 kpa
·Pressure of pipeline inlet:
O2,280kPa~600kPa/41Psi~87Psi, ≦180L/min
N2O,280kPa~600kPa/41Psi~87Psi, ≦180L/min
Air,280kPa~600kPa/41Psi~87Psi, ≦180L/min
·N2O cut off valve protection: when O2 is lost, N2O will be cut off.
Flow Meter
·6 tubes high precision flow meter with O2&N2O linkage device
·Gas scale:
O2: two tubes, 0~1L/min,1.1L/min~10L/min;
N2O: two tubes, 0~1L/min,1.1L/min~12L/min;
Air: two tubes, 0~1L/min,1.1L/min~15L/min;
·Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5.
·O2&N2O linkage device, to make insure O2 concentration is not lower than 25%
O2 flush: 35L/min~75L/min
Select Switch: select between C.G.O and CO2 Circle Absorber
C.G..O: alone used for open breath circuit
Selectatec Mounting bar: one unit place used for vaporizer, it is easy to replace vaporizer
Respiration System
·Absorber chamber:1000mL
·System leakage: when pressure is lower than 3kPa, ≦125L/min
·System compliance: ≦4ml/kPa
·Resistance of system:
Flow rate:30L/min﹤0.16kPa
Flow rate:60L/min﹤0.52kPa
·APL valve: 0.19~6kPa
Dimensions of the machine:
·Height:40cm(without CO2 circle absorber)/68cm (with CO2 circle absorber)
·Width:42cm(without CO2 circle absorber)/66cm (with CO2 circle absorber)
·Depth:20cm(without CO2 circle absorber)/23cm (with CO2 circle absorber)
·Weight: 21kg(with anesthesia vaporizer and without CO2 circle absorber)/ 24kg(with anesthesia vaporizer and with CO2 circle absorber)
·Gas source:
3 gas, O2,N2O,Air
·Gas source pressure gauge:
3 gas, O2,N2O,Air
·Output pressure of regulator: 280kpa~300kpa
·Security valve release pressure:≦600 kpa
·Pressure of pipeline inlet:
O2,280kPa~600kPa/41Psi~87Psi, ≦180L/min
N2O,280kPa~600kPa/41Psi~87Psi, ≦180L/min
Air,280kPa~600kPa/41Psi~87Psi, ≦180L/min
·N2O cut off valve protection: when O2 is lost, N2O will be cut off.
Flow Meter
·6 tubes high precision flow meter with O2&N2O linkage device
·Gas scale:
O2: two tubes, 0~1L/min,1.1L/min~10L/min;
N2O: two tubes, 0~1L/min,1.1L/min~12L/min;
Air: two tubes, 0~1L/min,1.1L/min~15L/min;
·Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5.
·O2&N2O linkage device, to make insure O2 concentration is not lower than 25%
O2 flush: 35L/min~75L/min
Select Switch: select between C.G.O and CO2 Circle Absorber
C.G..O: alone used for open breath circuit
Selectatec Mounting bar: one unit place used for vaporizer, it is easy to replace vaporizer
Respiration System
·Absorber chamber:1000mL
·System leakage: when pressure is lower than 3kPa, ≦125L/min
·System compliance: ≦4ml/kPa
·Resistance of system:
Flow rate:30L/min﹤0.16kPa
Flow rate:60L/min﹤0.52kPa
·APL valve: 0.19~6kPa
Dimensions of the machine:
·Height:40cm(without CO2 circle absorber)/68cm (with CO2 circle absorber)
·Width:42cm(without CO2 circle absorber)/66cm (with CO2 circle absorber)
·Depth:20cm(without CO2 circle absorber)/23cm (with CO2 circle absorber)
·Weight: 21kg(with anesthesia vaporizer and without CO2 circle absorber)/ 24kg(with anesthesia vaporizer and with CO2 circle absorber)